The fastest gradient cycle time is achieved with the innovative hardware technology. The total system delay volume has been reduced to less than 42 µL with the ultra-low volume (20 µL) mixer, optional loop injection kit with 5 µL loop and micro-volume pre-heater.
Because complex samples require high-sensitivity analysis, the front-end component for LC/MS(/MS) must decrease its background noise and carryover to zero. The SIL-30AC autosampler enables excellent LC/MS(/MS) analysis with near-zero carryover and, when necessary, can utilize its multi-rinse functionality, incorporating up to 4 different rinse solutions, for an even cleaner analysis.



| Desipramine | Amytriptyline | |
|---|---|---|
| ①+②+③+④ | Not Detected (<0.0042% = 1.0pg) | Not Detected (<0.0027% =0.68pg) |
| ①+②+③ | Not Detected (<0.0042% = 1.2pg) | Not Detected (<0.0027% =0.68pg) |
| ①+②* | Not Detected (<0.0042% = 1.2pg) | Not Detected (<0.0027% =0.68pg) |
| ①* | 0.01% | 0.01% |
| no rinse | 0.05% | 0.05% |

The SIL-30AC autosampler incorporates pre-treatment functionalities such as dilution, addition of an internal standard and pre-column derivatization schemes. These new features enable unattended operation for increased accuracy, precision and speed for high-throughput operation.



Green chromatography requires stable and reliable temperature control at a high temperature. The CTO-30A column oven provides precise temperature control up to 150ºC with a newly designed solvent pre-heater, a proprietary Intelligent Heat Balancer (IHB) and post-column solvent cooling. The solvent pre-heater and IHB ensure uniform column temperature even at the high flow rates achievable with Nexera.