Expandability

Premier UHPLC as a Front-end for Mass Spectrometers

The fastest gradient cycle time is achieved with the innovative hardware technology. The total system delay volume has been reduced to less than 42 µL with the ultra-low volume (20 µL) mixer, optional loop injection kit with 5 µL loop and micro-volume pre-heater.

Because complex samples require high-sensitivity analysis, the front-end component for LC/MS(/MS) must decrease its background noise and carryover to zero. The SIL-30AC autosampler enables excellent LC/MS(/MS) analysis with near-zero carryover and, when necessary, can utilize its multi-rinse functionality, incorporating up to 4 different rinse solutions, for an even cleaner analysis.

Nexera and LCMS-2020
Nexera and LCMS-2020

Simultaneous Analysis with Shimadzu LCMS-2020

Positive
  1.  1.  Atenolol
  2.  2.  Procaine
  3.  3.  Lidocaine
  4.  4.  Atropine
  5.  5.  Yohimbine
  6.  6.  Chlorpheniramine
  7.  7.  Propranolol
  8.  8.  Alprenolol
  9.  9.  Tetracaine
  10. 10. Diphenhydramine
  11. 11. Doxepin
  12. 12. Dipyridamol
  13. 13. Desipramine
  14. 14. Imipramine
  15. 15. Nortriptyline
  16. 16. Amitriptyline
  17. 17. Dibucaine
  18. 18. Verapamil
  19. 19. Reserpine
  20. 20. Carbamazepine
  21. 21. Isopropylantipyrine
  22. 22. Alprazolam
  23. 23. Trizolam
  24. 24. Cilostazol
  25. 25. Nifedipine
  26. 26. Diazepam
  27. 27. Warfarin
Negative
  1.  1.  Cefuroxime
  2.  2.  Chloramphenicol
  3.  3.  Nitrendipine
Column
: ODS column (2.1mm i.d. x 50mm, 1.8µm)
Mobile phase
: A: 0.1% formic acid in water
B: 0.1% formic acid in acetonitrile
Gradient
: B 3% → 95% (0.50 min)
Flow rate
: 1.8mL/min
Column temp.
: 50ºC
Detection
: ESI (+/-), LCMS-2020
Scan speed
: 15000 u/sec

Complete Reduction of Sample Carryover by Multi-Rinse Functionality

Complete Reduction of Sample Carryover by Multi-Rinse Functionality

[Analytical Conditions]

Column
: ODS column (2.1 mm i.d. x 50 mm, 2.2 µm)
Mobile phase
: A: 0.1% formic acid
B: acetonitrile
Flow rate
: 1.5 mL/min
Gradient
: B 25% → 50% (0.75 min) → 90% (0.76 min)
Column temp.
: 40ºC
Detection
: LCMS-2020, ESI (+)
Rinse solution
: R0: 0.1% formic acid in acetonitrile
: R1: 0.1% formic acid in methanol
Sample
: desipramine, amytriptyline (each 5 µg/mL)
Injection vol.
: 5 µL
  • ①: Needle Dip Rinse
  • ③: Internal Needle Rinse
  • ②: Needle Pump Rinse
  • ④: Injection Port Rinse
  Desipramine Amytriptyline
①+②+③+④ Not Detected (<0.0042% = 1.0pg) Not Detected (<0.0027% =0.68pg)
①+②+③ Not Detected (<0.0042% = 1.2pg) Not Detected (<0.0027% =0.68pg)
①+②* Not Detected (<0.0042% = 1.2pg) Not Detected (<0.0027% =0.68pg)
* 0.01% 0.01%
no rinse 0.05% 0.05%
*In case “①+②” or “①”, rinse solution for dip rinse is 0.1% formic acid in acetonitrile.

Automated Sample Pre-treatment

The SIL-30AC autosampler incorporates pre-treatment functionalities such as dilution, addition of an internal standard and pre-column derivatization schemes. These new features enable unattended operation for increased accuracy, precision and speed for high-throughput operation.

SIL-30AC Autosampler
SIL-30AC Autosampler

Analysis of o-phthal aldehyde using SIL-30AC automatic pretreatment function.

Column
: ODS column (3.0mm i.d. x 50mm, 2.2µm)
Mobile phase
: A: 20 mmol/L Potassium Phosphate
B: Acetonitrile
Flow rate
: 1.2 mL/min
Column temp.
: 40ºC
Detection
: Fluorescence at Ex 350 nm, Em 450 nm

Green Chromatography at 150ºC

Green chromatography requires stable and reliable temperature control at a high temperature. The CTO-30A column oven provides precise temperature control up to 150ºC with a newly designed solvent pre-heater, a proprietary Intelligent Heat Balancer (IHB) and post-column solvent cooling. The solvent pre-heater and IHB ensure uniform column temperature even at the high flow rates achievable with Nexera.

CTO-30A Column Oven
CTO-30A Column Oven

Comparison between Green and Conventional LC

Comparison between Green and Conventional LC

[Green LC]

Column
: Polymer-based ODS (3 mm i.d. x 150 mm, 3 µm)
Mobile phase
: Water only
Flow rate
: 0.5 mL/min
Column temp.
: 150ºC

[Conventional LC]

Column
: ODS column (4.6 mm i.d. x 150 mm, 5 µm)
Mobile phase
: Methanol/Water (30/70)
Flow rate
: 1.0 mL/min
Column temp.
: 40ºC