Product

Laboratory Instruments

GC-2010

Gas Chromatographs

GC-2010

Since introducing our first Gas Chromatograph over 50 years ago, Shimadzu has set the standard for GC power, speed and precision. Today, the GC-2010, in conjunction with GCsolution Software, is the most powerful gas chromatography analysis system available in the world, delivering faster analysis, reduced downtime, superior reproducibility, and excellent stability.

Features

Excellent Productivity (Highs-speed GC)
Demands for improved productivity and more efficiency have increased in recent years. The GC-2010 has been specially designed to address these demands by incorporating the following:

  • Third-generation Advanced Flow Control (AFC) – Supports high-speed analysis at a maximum pressure of 970 kPa and 1200 mL/minute maximum flow rate
  • High-power Oven – Enables cooling from 450° C to 50° C in less than six minutes
  • Rapid Detector Data Sampling Rate – Ensures sharp, accurate peaks

Together, the GC-2010 can double or even triple the productivity of conventional instruments, making it a leader in high-speed GC.

Superior Reproducibility
The GC-2010 is the ideal solution for all applications where reproducibility requirements are critical. Expect superb reproducibility – even for large vapor-volume solvents such as acetone or water. Our AFC features rapid response and temperature correction functions to precisely control pressure and flow. New compensation technology for ambient temperatures provides prolonged stability and the most repeatable retention times.

 

Constant Linear Velocity
Maintains optimum flow during temperature programming for the best chromatographic separation. Up to seven pressure ramps with either increasing or decreasing pressure gradients can be used to simplify complex separations.

Easy Operation
A number of features enable easy operation for both new and experienced users. These include:

  • Large LCD Screen (G window) – Shows graphically all of the operation of the GC, including the data results, temperature profile, and the chromatogram in progress
  • Self-diagnosis Functionality – Maximizes uptime of the GC system and provides documentation for validation purposes; self-diagnosis functions include septum and insert operating status, temperature sensor check, supplied gas pressure, DC voltage and AD converter.
  • Intuitive Operation with GCsolution (Assistant Bar, Data Explorer, Batch Table Wizard, and more)

Compact, Highly Sensitive Detectors
Ideal for a wide range of applications, all detectors (FID, TCD, ECD, FPD, FTD) are conveniently located under the top cover, and each detector incorporates an APC (Advanced Pressure Controller) that permits digital setting of the detector gas flow rate. A Flame Photometric Detector (FPD) provides 4 to 5 times more sensitivity than before, while the ECD delivers ultra-high sensitivity detection at 8fg/s.

Applications

GC-2010 Analysis of Biodiesel for Free and Total Glycerines by ASTM-6584
(pdf - 943 kb

Specifications

Column Oven
Temperature Range (Ambient + 4°C) ~ 450°C (using LCO2 gas: -50°C - 450°C)
Inner Dimensions and Volume280 (W) x 280 (H) x 175 (D) mm, 13.7L
Temperature accuracy±1% of the set temperature in Kelvin unit (calibration at 0.01°C increments)
Temperature Variation Coefficient 0.01°C/°C
Temperature Program StepsUp to 20 (cooling program possible)
Programmed rate setting range-250°C ~ 250°C/min
Total time for all steps9999.99 minutes max.

Injection Ports
Temperature RangeUp to 400°C
Heating Settings1°C steps
No. of units installed simultaneouslyUp to 3 units
Sample injection unit typesSplit/splitless injection is standard
Direct (without split) injection unit (option)
On-column/PTV (programmable temperature vaporizer)injection unit (option)

Flow Control Unit
Advanced Flow Controller (AFC)
Split/splitles injection mode and direct injection modePressure range: 0-970kPa
Programmable Steps7 (Negative pressure step is available)
Programmed Rate Setting Range-400 ~ 400kPa/min
Split Ratio Range0 - 9999.9
Mass Flow Rate Setting Range0 - 1200mL/min
Constant Linear Velocity Maintains optimum separation at all times

Available Optional Accessories
AOC-20i Auto Injector Direct injection, PTV/LVI, 1 - 250 uL syringes, ambient headspace
AOC-20s Auto Sampler150 1.5mL vials or 96 4mL vials and barcode reader
Leap Combi-Pal AutosamplerSPME, headspace and micro-titer plate sampling capability
PY2020D PyrolyzerSingle or double shot includes autosampler

.
Detectors
Four detectors can be installed simultaneously and individually temperature controlled. Detector gas is electronically controlled by APC (Advanced Pressure Control)
Detector Type(s)FID, TCD, ECD, FPD, FTD
Flame Ionization Detector (FID)
Temperature Range450°C max.
Minimum detected quantity3pgC/s (Dodecane)
Dynamic Range107
Thermal Conductivity Detector (TCD)
Temperature Range400°C max.
Dynamic Range105
Sensitivity20,000 mV - mL/mg (Decane)
Electron Capture Detector (ECD)
Temperature Range400°C max.
Minimum detected quantity8fg/s
Dynamic Range104
Flame Photometric Detector (FPD)
Temperature Range350°C max.
Dynamic RangeP: 104
S: 103
Minimum detected quantityP: 0.2pgP/s (tributyl phosphate)
S: 4pgS/s (dodecane thiol)
Flame Thermionic Detector (FTD)
(Two types, one for capillary and one for packed. The specifications are the same.)
Temperature Range450°C max.
Dynamic RangeN: 104
P: 104
Minimum detected quantityN: 0.2pgN/s (azobenzene)
P: 0.03pgP/s (malathion)
Maximum of 2 detectors to keep instrument performance

Dimensions, Weight, Power Requirements (GC main unit)
Dimensions515 (W) x 440 (H) x 530 (D) mm
Weight30 kg (FID model)
Power RequirementsAC100V/120V 230V
1800VA (Normal oven type) or
2600VA (High power oven type), 50/60Hz