Prominence UFLC
Liquid Chromatographs
Good repeatability is a strong feature of an HPLC, and it must not be sacrificed by high speed. Prominence UFLC employs a proven micro-stroke solvent delivery mechanism (flow rate resolution 3.7nL/min., 10uL/stroke) with improved firmware so that ultra-fast gradient formation occurs. The Prominence Autosampler, with its highly sealed measuring device and high-precision valves, achieves excellent repeatability for fast LC.
Prominence UFLC provides good repeatability even in small volume sample injection.
Repeatability in small-volume injection and ultra-fast analysis
In general high-speed LC, repeatability in a small sample volume is often regarded as a problem. The figure below shows a chromatogram by Prominence UFLC for a 2uL injection and one-minute gradient analysis. Repeatability of less than 0.2% RSD is achieved.
Column: Shim-pack XR-ODS (50mmL. x 3mm i.d.), Temperature: 40°C. Mobile Phase: Water/Acetonitrile gradient elution, Flow Rate: 1.5mL/min, Detection: 245nm, Sample: Alkylphenones, Injection Volume: 2uL
 |
| Peaks |
Retention time |
Peak area |
| Average |
%RSD |
Average |
%RSD |
| 1. |
Acetophenone |
0.377 |
0.076 |
49018 |
0.193 |
| 2. |
Propiophenone |
0.493 |
0.072 |
47260 |
0.167 |
| 3. |
Butyrophenone |
0.587 |
0.070 |
49199 |
0.055 |
| 4. |
Valerophenone |
0.677 |
0.065 |
45277
|
0.138 |
| 5. |
Hexanophenone |
0.763 |
0.082 |
46613 |
0.186 |
| 6. |
Heptanophenone |
0.842 |
0.087 |
43975 |
0.205 |
| 7. |
Octanophenone |
0.914 |
0.080 |
48978 |
0.200 |
|
|
For precise equipment control, required by analytical validation processes and quality control tests, and for minimizing the risk of a system malfunction interrupting an analysis, the most important factor is the high-pressure valve on the autosampler. The Prominence UFLC uses Shimadzu’s own proprietary high-precision 6-port valve, which achieves 0.1% or less (RSD) area repeatability in 100,000 cycle endurance tests*. In addition, for LC-MS analyses, where reduction of sample carryover has currently gained a great deal of attention, Prominence UFLC successfully reduces carryover based on the principle of inhibiting the adsorption of samples in the first place. * Based on Shimadzu endurance test parameters for a 6-port valve.
Repeatability for 4 Alkyl Phenons After Finishing 100,000 Cycle Endurance Test
 |
| Peaks |
Retention time |
Area |
| Average |
%RSD |
Average |
%RSD |
| 1. |
Acetophenone |
0.916 |
0.089 |
312670 |
0.059 |
| 2. |
Propiophenone |
1.398 |
0.058 |
315739 |
0.045 |
| 3. |
Butyrophenone |
2.111 |
0.030 |
336428 |
0.063 |
| 4. |
Valerophenone |
3.341 |
0.031 |
296609 |
0.040 |
|
|
<< Go Back Next UFLC Page >>