AOC Series

  AOC-20i AOC-20s
Sample Injection Method Liquid sample injection via special microsyringe
Sample Volume 0.1 ~ 8.0 µl, 0.1µl steps (using 10ul syringe)
0.5 ~ 40 µl, 0.5 µl steps (using 50ul syringe)
2.5 ~ 200 µl, 2.5 µl steps (using 250ul syringe)
Number of Samples 6 vials (with option, 12 possible) 1.5ml sample vials - 150 vials
4ml sample vials- 96 vials
Sample Vials Glass construction, 1.5ml, 4ml, screw top, PTFE-coated septum
Rinse Solvent Vials Glass construction, 4ml, screw top, PTFE-coated septum
Number of Sample Injections 1 ~ 99 injections per sample
Syringe Speed 2 Modes: Fast and Slow
Plunger Speed 3 Modes: Fast, Medium and Slow
Wait Time 0 ~ 99 sec following sample aspiration (in 0.1 sec steps)
Type of Sample Injection 3 Modes: Traditional, Solvent Flush and Solvent Flush with a Second Solvent
Injection Volume Linearity ±0.5% (injection volume 1 ~ 5 µl, sample n-C12
Cross Contamination Less than 10-4 (as determins with a 1% diphenyl in hexane using 4 solvent rinses)
Priority Sample A priority sample injection can be injected during a sample sequence, the the sequence can be resumed
Dual Injection System Available on GC-17A platform only One set of AOC-20s can feed sample vials to two sets of AOC-20i
Sample Cooling/Heating 0 ~ 60°C, with optional cooling rack connected to a general laboratory circulating water bath
External Control Includes optional optical link or RS232C interface
Dimensions, weight 126 (W) x 380 (H) x 78 (D) mm, 2.5 kg 320 (W) x 135 (H), 2.4 kg
Power Supply 260 (W) x 70 (H) x 420 (D) mm, 2.8kg
Power requirements AC100/115/220V±10% 50/60HZ

For Research Use Only. Not for use in diagnostic procedures.