Nexera X2 UHPLC System

Nexera X2 UHPLC System

The Shimadzu Nexera X2 ultra high-performance liquid chromatograph (UHPLC) is the most advanced chromatographic system available today. This UHPLC system offers unsurpassed chromatographic efficiency, sensitivity, resolution, stability, and reliability. The Nexera X2 is a truly flexible design that expands to tackle a broad range of UHPLC applications.

Specifications

Hardware Specifications (Limited)

Here are limited specifications for the typical components used in the Nexera XR system.

Detailed specifications are located within the component's pages.

System Controllers
Degassers
Pumps
Autosamplers
Column Ovens
Detectors

Pump

 
LC-30AD
Degassing Unit
3 or 5 Lines (Volume 400 μL)
Flow rate range
0.0001 to 10 mL/min
Flow rate precision
No more than 0.06% RSD or 0.02 min SD, whichever is greater
Gradient type
High-pressure / low-pressure mixing gradient
Gradient / concentration precision
±0.5% (specified conditions)
Maximum pressure
130 MPa (0.0001 to 3 mL/min)
80 MPa (3.0001 to 5 mL/min)
22Mpa (5.0001 to 10 mL/min)

Autosampler

 
SIL-30A/C
SIL-30ACMP
Injection method
Needle-in-flowpath
Needle-in-flowpath
Injection volume accuracy    
±1% (at 50 μL injectiong, n = 10)
±1% (at 50 μL injectiong, n = 10)
Injection volume range
0.1 to 50 μL (Option: 0.1 to 100 μL)
0.1 to 50 μL (Option: 0.1 to 100 μL)
Injection volume reproducibility
RSD: 0.3% maximum (at 10 μL) RSD: 0.2% maximum (at 10 μL)
Samples capacity Sample plate: 175 (1 mL), 105 (1.5 mL), 50 (4 mL), 2 (MTP/DWP plates)
Control plate: 10 (1.5 mL)
Sample plate: 324 (1.5 mL), 576 (96 well MTP/DWP Plates), 2304 (384 well MTP/DWP plates)
Control plate: 10 (1.5 mL)
Sample cooler
Block cooling/heating, used together with defumidifying function
4 to 40°C
Block cooling/heating, used together with defumidifying function
4 to 40°C

Oven

 
CTO-30A
Oven Capacity
Up to 2 - 150 mm columns
Temperature control range
Room temperature ± 5 to 150°C, Setting range 4 to 150°C

PDA Detector

 
SPD-M30A
Wavelength Range
190 to 800 nm
Noise level
±3 × 10?6 AU, (250 nm, Specified condition)
Flow cell
10 μL (10 mm), 8 MPa
Option cell
High-Sensitivity: 8 μL (10 mm)

For Research Use Only. Not for use in diagnostic procedures.