Prominence HPLC System

Prominence HPLC

The Shimadzu Prominence high-performance liquid chromatograph (HPLC) achieves an exceptional level of performance, reliability, and economy. Prominence HPLC covers a wide variety of flow rates and application goals.

Specifications

Hardware Specifications (Limited)

Here are limited specifications for the typical components used in the Nexera XR system.

Detailed specifications are located within the component's pages.

System Controllers
Degassers
Pumps
Autosamplers
Column Ovens
Detectors

 
 

Pump

  LC-20AD LC-20AT LC-20AB
Degassing Unit 3 or 5 Lines (Volume 400 μL)
Pulsation 0.1 MPa (for water at 1.0 mL/min. and 7MPa) 0.3 MPa (for water at 1.0 mL/min. and 7MPa) 0.1 MPa (for water at 1.0 mL/min. and 7MPa)
Flow rate range 0.0001 to 10.0000 mL/min. 0.001 to 10.000 mL/min. 0.0001 to 10.0000 mL/min.
Flow rate precision No more than 0.06% RSD or 0.02 min SD, whichever is greater
Gradient type High-pressure / low-pressure mixing gradient High Pressure mixing
Gradient / concentration precision ±0.5% (specified conditions) 1.0% (specified conditions) ±0.5% (specified conditions)
Maximum pressure 40 MPa
 
 

Autosampler

  SIL-20A/C HT
Injection method Needle-in-flowpath
Injection volume accuracy ±1% (at 100 μL injection, n = 10)
Injection volume range 0.1 to 100 μL (Option: 0.1 to 50 μL, 1 to 500 μL, 1 to 2,000 μL)
Injection volume reproducibility RSD: 0.3% maximum (at 10 μL injection)
Samples capacity Sample plate: 175 (1 mL), 105 (1.5 mL), 50 (4 mL), 2 (MTP/DWP plates)
Control plate: 10 (1.5 mL)
Sample cooler Block cooling/heating, used together with defumidifying function
4 to 40°C
*No sampler cooler included in SIL-20AHT model
 
 

Oven

  CTO-20A/C
Oven Capacity
W220 × D95 × H365 mm
Temperature control range Room temperature ± 10 to 85°C, Setting range 4 to 85°C
 
 

UV Detector

  SPD-20AV
Wavelength Range
190 to 900 nm (with optional tungsten lamp)
Noise level ±2.5 × 10?6 AU, (250 nm, Specified condition)
Flow cell 12 μL (10 mm), 12 MPa
Option cell High-Sensitivity: 8 μL (10 mm)
Semi-micro: 2.5 μL (5mm)
 
 

PDA Detector

  SPD-M20A
Wavelength Range
190 to 800 nm
Noise level
±6 × 10?6 AU, (250 nm, Specified condition)
Flow cell
10 μL (10 mm), 8 MPa
Option cell
High-Sensitivity: 8 μL (10 mm)

For Research Use Only. Not for use in diagnostic procedures.