Main Unit PPSQ-51A PPSQ-53A
Reaction method Edman degradation
Reaction time 46.5 min/cycle 48 min/cycle
Number of reactors 1 3
Sample immobilization method Glass fiber disk (8 mm dia.) or PVDF membrane
Reactor temperature control range 10 °C above room temperature to 60 °C
Converter temperature control range 10 °C above room temperature to 70 °C
Number of reagents/solvents 7
Reagent/solvent delivery method Nitrogen gas pressure
Dimensions W510 x D500 x H540 mm
Weight 43 kg 45 kg

PPSQ-51A
Single Reactor


Single Reactor

Offers highly cost-effective performance.

PPSQ-53A
Triple Reactor
(Supports consecutive analysis of 3 samples.)

Triple Reactor

Using a triple reactor makes it possible to save time and effort, and gives greater freedom in the formulation of analysis programs.

Standard Configurations

Isocratic System

Configuration PPSQ-51A/53A (main unit)
LC-40D
SPD-M30A
CTO-40C

Gradient System

Configuration PPSQ-51A/53A (main unit)
LC-40D (two)
SPD-M30A
CTO-40C
DGU-403

Other Items Provided by User

Installation space Desktop: W1,600 x D600 x H800 mm min.
Weight: Approx. 100 kg
Power supply 120-230 VAC, 50/60 Hz, 1,500 VA max.
Nitrogen gas Purity: 99.9999% min.
A cylinder pressure regulator and gas tubing (10 m) are provided as standard accessories.
Exhaust equipment In order to provide an exhaust for gases produced by waste liquids, an exhaust tube must either be connected to exhaust equipment or vented outside. An exhaust tube (20 m) is provided as a standard accessory.

Note: The required installation space and power supply may change according to the PC and display used.

 

Installation Example

Isocratic System

Isocratic System

Gradient System

Gradient System

SPD-M30A

Detector

Wavelength setting range

190 to 700 nm

Dimensions

W260 x D500 x H140 mm

Weight

12 kg

CTO-20A

Oven

Temperature control range

10 °C above room temperature to 60 °C

Dimensions

W260 x D420 x H415 mm

Weight

20 kg

PPSQ-51A System

Standard Configurations

Configuration

PPSQ-51A (main unit)
LC-20AT
SPD-M30A
CTO-20A

PPSQ-53A System

Standard Configurations

Configuration

PPSQ-53A (main unit)
LC-20AT
SPD-M30A
CTO-20A

Other Items Provided by User

Installation space

Desktop: W1,600 x D600 x H800 mm min.
Weight: Approx. 100 kg

Power supply

120-230 VAC, 50/60 Hz, 1,500 VA max.

Nitrogen gas

Purity: 99.9999% min.
A cylinder pressure regulator and gas tubing (10 m) are provided as standard accessories.

Exhaust equipment

In order to provide an exhaust for gases produced by waste liquids, an exhaust tube must either be connected to exhaust equipment or vented outside. An exhaust tube (20 m) is provided as a standard accessory.

Note: The required installation space and power supply may change according to the PC and display used.

Read more
Show Less

PPSQ-51A/53A

Main Unit

Reaction method

Edman degradation

Reaction time

PPSQ-51A: 46.5 min/cycle
PPSQ-53A: 48 min/cycle

Number of reactors

PPSQ-51A: 1
PPSQ-53A: 3

Sample immobilization method

Glass fiber disk (8 mm dia.) or PVDF membrane

Reactor temperature control range

10 °C above room temperature to 60 °C

Converter temperature control range

10 °C above room temperature to 70 °C

Number of reagents/solvents

7

Reagent/solvent delivery method

Nitrogen gas pressure

Dimensions

W510 x D500 x H540 mm

Weight

PPSQ-51A: 43 kg
PPSQ-53A: 45 kg

Read more
Show Less

LC-20AD

Solvent Delivery Module

Flow rate setting range

0.0001 to 10 mL/min

Dimensions

W260 x D420 x H140 mm

Weight

10 kg

SPD-M30A

Detector

Wavelength setting range

190 to 700 nm

Dimensions

W260 x D500 x H140 mm

Weight

12 kg

CTO-20AC

Oven

Temperature control range

Room temperature - 10 °C to 60 °C

Dimensions

W260 x D420 x H415 mm

Weight

23 kg

Control PC

OS

Windows 10 Pro 64 bit

Note: Windows is a registered trademark of Microsoft Corporation in the U.S. and other countries.

PPSQ-51A/53A

Main Unit

Reaction method

Edman degradation

Reaction time

PPSQ-51A: 46.5 min/cycle
PPSQ-53A: 48 min/cycle

Number of reactors

PPSQ-51A: 1
PPSQ-53A: 3

Sample immobilization method

Glass fiber disk (8 mm dia.) or PVDF membrane

Reactor temperature control range

10 °C above room temperature to 60 °C

Converter temperature control range

10 °C above room temperature to 70 °C

Number of reagents/solvents

7

Reagent/solvent delivery method

Nitrogen gas pressure

Dimensions

W510 x D500 x H540 mm

Weight

PPSQ-51A: 43 kg
PPSQ-53A: 45 kg

Read more
Show Less

LC-20AT

Solvent Delivery Module

Flow rate setting range

0.001 to 10 mL/min

Dimensions

W260 x D420 x H140 mm

Weight

11 kg

PPSQ-51A System

Standard Configurations

Configuration

PPSQ-51A (main unit)
LC-20AD (two)
SPD-M30A
CTO-20AC
DGU-20A3R

PPSQ-53A System

Standard Configurations

Configuration

PPSQ-53A (main unit)
LC-20AD (two)
SPD-M30A
CTO-20AC
DGU-20A3R

Other Items Provided by User

Installation space

Desktop: W1,600 x D600 x H800 mm min.
Weight: Approx. 100 kg

Power supply

120-230 VAC, 50/60 Hz, 1,500 VA max.

Nitrogen gas

Purity: 99.9999% min.
A cylinder pressure regulator and gas tubing (10 m) are provided as standard accessories.

Exhaust equipment

In order to provide an exhaust for gases produced by waste liquids, an exhaust tube must either be connected to exhaust equipment or vented outside. An exhaust tube (20 m) is provided as a standard accessory.

Note: The required installation space and power supply may change according to the PC and display used.

Read more
Show Less

Control PC

OS

Windows 10 Pro 64 bit

Note: Windows is a registered trademark of Microsoft Corporation in the U.S. and other countries.

{"title":"Specification Sheet Downloads","description":"Download the latest specification sheets.","source":"product","key":3414,"max":"30","filter_types":["spec_sheet"],"link_title":"View other Downloads","link_url":""} {"title":"Downloads","description":"Download the latest brochure.","source":"product","key":3414,"max":"30","filter_types":["brochures"],"link_title":"View other Downloads","link_url":"","pdf_links":[]}