PPSQ-51A/53A Protein Sequencers - Specs
Confident Protein/Peptide Sequencing using Automated Edman Degradation
| Main Unit | PPSQ-51A | PPSQ-53A |
| Reaction method | Edman degradation | |
| Reaction time | 46.5 min/cycle | 48 min/cycle |
| Number of reactors | 1 | 3 |
| Sample immobilization method | Glass fiber disk (8 mm dia.) or PVDF membrane | |
| Reactor temperature control range | 10 °C above room temperature to 60 °C | |
| Converter temperature control range | 10 °C above room temperature to 70 °C | |
| Number of reagents/solvents | 7 | |
| Reagent/solvent delivery method | Nitrogen gas pressure | |
| Dimensions | W510 x D500 x H540 mm | |
| Weight | 43 kg | 45 kg |
PPSQ-51A
Single Reactor
Offers highly cost-effective performance.
PPSQ-53A
Triple Reactor
(Supports consecutive analysis of 3 samples.)
Using a triple reactor makes it possible to save time and effort, and gives greater freedom in the formulation of analysis programs.
Other Items Provided by User
| Installation space | Desktop: W1,600 x D600 x H800 mm min. |
| Weight: Approx. 100 kg | |
| Power supply | 120-230 VAC, 50/60 Hz, 1,500 VA max. |
| Nitrogen gas | Purity: 99.9999% min. |
| A cylinder pressure regulator and gas tubing (10 m) are provided as standard accessories. | |
| Exhaust equipment | In order to provide an exhaust for gases produced by waste liquids, an exhaust tube must either be connected to exhaust equipment or vented outside. An exhaust tube (20 m) is provided as a standard accessory. |
Note: The required installation space and power supply may change according to the PC and display used.
Installation Example
Isocratic System

Gradient System

PPSQ-51A/53A
Main Unit
| Reaction method |
Edman degradation |
|---|---|
| Reaction time |
PPSQ-51A: 46.5 min/cycle |
| Number of reactors |
PPSQ-51A: 1 |
| Sample immobilization method |
Glass fiber disk (8 mm dia.) or PVDF membrane |
| Reactor temperature control range |
10 °C above room temperature to 60 °C |
| Converter temperature control range |
10 °C above room temperature to 70 °C |
| Number of reagents/solvents |
7 |
| Reagent/solvent delivery method |
Nitrogen gas pressure |
| Dimensions |
W510 x D500 x H540 mm |
| Weight |
PPSQ-51A: 43 kg |
LC-20AT
Solvent Delivery Module
| Flow rate setting range |
0.001 to 10 mL/min |
|---|---|
| Dimensions |
W260 x D420 x H140 mm |
| Weight |
11 kg |
SPD-M30A
Detector
| Wavelength setting range |
190 to 700 nm |
|---|---|
| Dimensions |
W260 x D500 x H140 mm |
| Weight |
12 kg |
CTO-20A
Oven
| Temperature control range |
10 °C above room temperature to 60 °C |
|---|---|
| Dimensions |
W260 x D420 x H415 mm |
| Weight |
20 kg |
Other Items Provided by User
| Installation space |
Desktop: W1,600 x D600 x H800 mm min. |
|---|---|
| Power supply |
120-230 VAC, 50/60 Hz, 1,500 VA max. |
| Nitrogen gas |
Purity: 99.9999% min. |
| Exhaust equipment |
In order to provide an exhaust for gases produced by waste liquids, an exhaust tube must either be connected to exhaust equipment or vented outside. An exhaust tube (20 m) is provided as a standard accessory. |
Note: The required installation space and power supply may change according to the PC and display used.
PPSQ-51A/53A
Main Unit
| Reaction method |
Edman degradation |
|---|---|
| Reaction time |
PPSQ-51A: 46.5 min/cycle |
| Number of reactors |
PPSQ-51A: 1 |
| Sample immobilization method |
Glass fiber disk (8 mm dia.) or PVDF membrane |
| Reactor temperature control range |
10 °C above room temperature to 60 °C |
| Converter temperature control range |
10 °C above room temperature to 70 °C |
| Number of reagents/solvents |
7 |
| Reagent/solvent delivery method |
Nitrogen gas pressure |
| Dimensions |
W510 x D500 x H540 mm |
| Weight |
PPSQ-51A: 43 kg |
LC-20AD
Solvent Delivery Module
| Flow rate setting range |
0.0001 to 10 mL/min |
|---|---|
| Dimensions |
W260 x D420 x H140 mm |
| Weight |
10 kg |
SPD-M30A
Detector
| Wavelength setting range |
190 to 700 nm |
|---|---|
| Dimensions |
W260 x D500 x H140 mm |
| Weight |
12 kg |
CTO-20AC
Oven
| Temperature control range |
Room temperature - 10 °C to 60 °C |
|---|---|
| Dimensions |
W260 x D420 x H415 mm |
| Weight |
23 kg |
Control PC
| OS |
Windows 10 Pro 64 bit |
|---|
Note: Windows is a registered trademark of Microsoft Corporation in the U.S. and other countries.
Other Items Provided by User
| Installation space |
Desktop: W1,600 x D600 x H800 mm min. |
|---|---|
| Power supply |
120-230 VAC, 50/60 Hz, 1,500 VA max. |
| Nitrogen gas |
Purity: 99.9999% min. |
| Exhaust equipment |
In order to provide an exhaust for gases produced by waste liquids, an exhaust tube must either be connected to exhaust equipment or vented outside. An exhaust tube (20 m) is provided as a standard accessory. |
Note: The required installation space and power supply may change according to the PC and display used.