HITS-X Series - Specs

High-Speed Impact Testing Machines

High Speed Puncture Impact Testing Machine HITS-PX

Main Unit, Controller and Software

Product Number

346-72457

Safety Devices

Door open/close interlock switch Safety mat interlock switch Two-switch start operation Start timer, etc.

Dimensions (Main Testing Unit)

Approx. W1,100 × D900 × H2850 mm

Air Supply Requirements

0.6 to 0.7 Mpa

Power Supply Requirements

200 V 3-phase 6 kVA, 100 V single-phase 1.5 kVA

PC Environment Required for Operation*2

Compatible OS: Windows 10 (Japanese, English) / Memory capacity: 4 GB or more /
HDD capacity: 120 GB or more / Display resolution and colors: 1024 × 768, 65,535 colors /
Other required peripheral equipment: CD-ROM / Expansion bus: One full-size PCI bus empty slot /
Communications: RS-232C communication port (required for communication with 4875 controllers)

Software

High-speed impact testing software

Controller

Model 4870 controller (specialized for high-speed impact testing)

Hydraulic Pressure Supply

Model AF-7H, 7 L /min, air cooled, installed below main unit

Impact Test Force

10 kN

Specimen Holder

Pneumtic clamping

AD Converter

Sampling rate: Max. 2 MHz with 12-bit resolution

Displacement Amplifier

Range: 10%, 20%, 50% or 100% of 150 mm / Accuracy: Within 1.0% of full scale Response frequency: DC- 10 kHz (-3 dB)

Force Amplifier

Range: 20%, 50% or 100% of load detector rating / Accuracy: Within 1.0% of full scale Response frequency: DC- 100 kHz (-3 dB)

Piston stroke

300 mm

Range Of Speed Settings

1 m/s to 20 m/s

Maximum Speed

20 m/s

Weight (Main Testing Unit)

Approx. 1,500 kg

Read more
Show Less

HITS-X Series

Impact Test Force

10 kN

Speed setting range

1 to 20 m/s

Piston stroke

300 mm

Controller

Controller 4870 (dedicated controller for high-speed impact testing)

Software

TRAPEZIUM HITS high-speed impact testing software

Read more
Show Less

High Speed Tensile Testing Machine HITS-TX

Main Unit, Controller and Software

Product Number

346-72545

Safety Devices

Door open/close interlock switch Safety mat interlock switch Two-switch start operation Start timer, etc.

Dimensions (Main Testing Unit)

Approx. W1,100 × D700 × H2,850 mm

Air Supply Requirements

Not required

Power Supply Requirements

200 V 3-phase 6 kVA, 100 V single-phase 1.5 kVA

PC Environment Required for Operation*2

Compatible OS: Windows 10 (Japanese, English) / Memory capacity: 4 GB or more /
HDD capacity: 120 GB or more / Display resolution and colors: 1024 × 768, 65,535 colors /
Other required peripheral equipment: CD-ROM / Expansion bus: One full-size PCI bus empty slot /
Communications: RS-232C communication port (required for communication with 4875 controllers)

Software

High-speed impact testing software

Controller

Model 4870 controller (specialized for high-speed impact testing)

Hydraulic Pressure Supply

Model AF-7H, 7 L /min, air cooled, installed below main unit

Impact Test Force

10 kN

Acceleration Jig

Tapered acceleration mechanism

AD Converter

Sampling rate: Max. 2 MHz with 12-bit resolution

Displacement Amplifier

Range: 10%, 20%, 50% or 100% of 150 mm / Accuracy: Within 1.0% of full scale Response frequency: DC- 10 kHz (-3 dB)

Force Amplifier

Range: 20%, 50% or 100% of load detector rating / Accuracy: Within 1.0% of full scale Response frequency: DC- 100 kHz (-3 dB)

Piston stroke

300 mm

Range Of Speed Settings

0.0001 m/s to 20 m/s*1

Maximum Speed

20 m/s

Weight (Main Testing Unit)

Approx. 1,500 kg

*1: Duration of one test must be within five minutes.
*2: The computer and operating system are not included in the standard system configuration. Please acquire them separately.

Read more
Show Less
{"title":"Specification Sheet Downloads","description":"Download the latest specification sheets.","source":"product","key":3472,"max":"30","filter_types":["spec_sheet"],"link_title":"View other Downloads","link_url":""} {"title":"Downloads","description":"Download the latest brochure.","source":"product","key":3472,"max":"30","filter_types":["brochures"],"link_title":"View other Downloads","link_url":"","pdf_links":[]}