Troubleshooting Peak Problems

Nexera Virtual Advisor

Symptom Cause Solution
  • Retention time fluctuation
Varying mobile phase composition Use a fresh mobile phase preparation and check the composition of the new mobile phase.
Retention time fluctuation Unstable flow rate Examine the pump function and take corrective action as necessary.
Retention time fluctuation Fluctuating ambient temperature Relocate the instrument or protect the current location from excessive environmental effects. Use a column oven.
Retention time fluctuation Column performance has deteriorated Examine the column oven function and take corrective action as necessary. Replace the column.
Retention time fluctuation Check valve is not working properly Clean the check valve in an ultrasonic bath or replace as necessary. Pump 2-propanol through the check valve to clean it.
Retention time fluctuation Blockage present in the needle or plumbing Clean the high-pressure valve with reverse cleaning of the flow lines. Remove the blockage with reverse cleaning of the flow lines.
  • No peaks/very small peaks
No plate in the specified rack position Insert a plate with samples in the specified rack position.
  • No peaks/very small peaks
Needle contacts the bottom of the plate or misses the target Determine the stack application (microtiter plate or deep-well microfilter plate) and set the needle stroke accordingly. Perform rack teaching for the SIL-20A/20AC, if applicable.
  • No peaks/very small peaks
Flow line blocked with plate material particles  Remove the blockage with reverse cleaning of the flow lines.
  • No peaks
No mobile phase flow Examine the pump function and take corrective action as necessary.
  • No peaks
Insufficient sample in vials Add more sample to the vials.
  • No peaks
Detector is not connected or is faulty Examine the detector connections and take corrective action as necessary.
No peaks Column performance has deteriorated Replace the column.
No peaks Blockage in the injection flow lines Check and correct the content of the user-written program. Clear or replace any blocked plumbing.
  • OVER is displayed
Leak - flow lines Find and repair the leak.
  • OVER is displayed
Dead volume between the flow line connections Examine flow line connections. Reconfigure the connections to eliminate any dead volume.
  • OVER is displayed
Connections between the pump and column are reversed Reconfigure the plumbing connections.
  • OVER is displayed
Column performance has deteriorated Replace the column.
  • OVER is displayed
Recorder pen is below the baseline Press [Zero] to return the pen to the the baseline. Remove the blockage with reverse cleaning of the flow lines. Replace the needle or the plumbing.
  • Poor reproducibility
Varying mobile phase composition Examine the pump functionality and take corrective action as necessary. Use a fresh mobile phase preparation and check the composition of the new mobile phase.
  • Poor reproducibility
Needle seal has deteriorated Replace the needle seal.
  • Poor reproducibility
Leak - flow lines Find and repair the leak.
  • Poor reproducibility
Insufficient flow line rinse or no rinse solution Increase the RINSE VOL settings or replenish the rinse solution. Perform the PUMP HEAD FLUSH operation. PURGE or RINSE the flow lines.
  • Poor reproducibility
Fluctuating ambient temperature Install unit in a room with minimal temperature variation.
  • Poor reproducibility
Column performance has deteriorated  Replace column.