Troubleshooting Valve Problems

Nexera Virtual Advisor

Symptom Cause Solution
  • Pressure too high - valve inlet
Valve position is misaligned  Contact your Shimadzu representative
Pressure too high - valve inlet Flow line blockage - valve Clean or replace any blocked flow lines
  • Position buttons will not change the valve position
Local mode is not enabled Press the Local/Remote key to enable local mode.
  • Leak - valve
Rotor and stator seal are worn Replace the rotor and stator as necessary.
  • Leak - valve
Pressure is set too high Set the pump PMAX to 34.3 MPa.
  • Leak - valve
Flow line connections are loose Tighten or replace the male nuts and ferrules of the flow line connections.
  • Leak - valve
Flow line connections are faulty Tighten or replace the male nuts and ferrules of the flow line connections.
  • Contact input does not move the valve position
Remote mode is not enabled Press the Local/Remote key to enable remote mode.
  • Contact input does not move the valve position
Event cable is improperly connected Connect the event cable properly.
  • Contact input does not move the valve position
Event cable is cut or broken Replace the event cable.
  • Power will not turn ON
Power supply specifications have not been set Use a power supply that meets the instrument specifications.
  • Power will not turn ON
Power cable is unplugged Properly connect the power cable to the power supply.
  • Power will not turn ON
Power cable is cut or broken Replace the power cable with a new one.
  • Power will not turn ON
Fuse is blown  Replace the fuses. If fuses blow out frequently, contact your Shimadzu representative.