USF-2000A - Specs

Ultrasonic Fatigue Testing System

USF-2000A

Test frequency

20 kHz ± 500 Hz

Test stress

180 to 900 MPa (in the case of a steel circular taper specimen)

Stress ratio

-1

USF-2000A

Test frequency

20 kHz ±500 Hz (recommended test range: 20 kHz ±30 Hz)
Note: The test frequency is determined from the resonance frequency of the sample.

Vibration at horn tip

±10 to ±50 μm

Test stress range

Stress given ±10 to ±50 μm displacement of sample
Note: Stress values depend on sample shape and physical property values.

Stress ratio

-1

Testable materials

Materials that can be resonated at 20 kHz and generate minimal heat during resonance
Example: High-strength steel, duralumin, titanium alloy, aluminum, etc.

Not-testable materials

- Materials that cannot resonate at 20 kHz
- Materials for which samples are difficult to attach
- Materials that generate significant heat during resonance at 20 kHz, due to friction
Examples: Resins, ceramics, etc.

Power Requirements

3-phase 200 V: 2 kVA (air compressor) , 1-phase 200V: 3.5 kVA (ultrasonic fatigue testing system) , 1-phase 100V: 1 kVA (computer, displacement logger, air dryer, etc.)

Standard contents

USF-2000A Ultrasonic Fatigue Testing System main unit (including table), ultrasonic resonance system, control computer,ultrasonic testing control and measurement software, and cooling unit (air dryer and compressed air lines)
Note: Air compressor for cooling is not included.

Required optional products

Air compressor (for regions with 50 Hz or 60 Hz power) 3-phase 200 V: 2 kVA
Displacement measuring system (eddy current displacement gauge with 0.5 μm resolution)
Note: A high-speed data logger or digital oscilloscope is required separately for reading voltages output from the displacement gauge.
Displacement gauge calibrator (CDE-25 C1 high-performance micrometer)

Note: Systems can be selected without an air compressor in cases where the customer will supply the compressed air. A 150 L/m flow rate of compressed air at a minimum 0.2 MPa is required.

Read more
Show Less
{"title":"Specification Sheet Downloads","description":"Download the latest specification sheets.","source":"product","key":3470,"max":"30","filter_types":["spec_sheet"],"link_title":"View other Downloads","link_url":""} {"title":"Downloads","description":"Download the latest brochure.","source":"product","key":3470,"max":"30","filter_types":["brochures"],"link_title":"View other Downloads","link_url":"","pdf_links":[]}