Constant-Temperature Measurement

Measurement of ALP Enzyme Activity

Measurement of ALP Enzyme Activity

The following accessories are used for biological component measurement, enzymatic activity measurement and reaction rate analysis for which the sample must be kept at a constant temperature.

Instrument: UV-2600 + CPS-100
Substrate: 0.1 mM
Enzyme: 0 - 0.05U
Temperature: 37°C
Detection Wavelength: 420 nm

CPS-100 6-Cell Thermoelectrically Temperature-Controlled Cell Positioner

CPS-100 6-Cell Thermoelectrically Temperature-Controlled Cell Positioner

Sample automation allows for switching between six cells simultaneously. Enables accurate temperature control without the need for circulation of cooling water.

  • For UVmini-1240 series, UV-1800/2600/2700/3600 Series
  •  Square cells are not included in the standard package. Please purchase separately.
  •  When used with the UV-1800 or UV-2600/2700, a USB adapter CPS is required.

TCC-100 Thermoelectrically Temperature Controlled Cell Holder

TCC-100 Thermoelectrically Temperature Controlled Cell Holder

An easy-to-use constant-temperature cell holder that uses a Peltier element for electronic cooling. It does not require an external thermostatic bath or water for cooling.

  • For UVmini-1240 series, UV-1800/2600/2700/3600 Series
  •  Square cells are not included in the standard package. Please purchase separately. 
  

Constant-Temperature Cell Holder

Constant-Temperature Cell Holder

This holder holds the sample cell at the desired, uniform temperature by circulating constant-temperature water. It accepts a pair of 10 mm square cells.

  • For UVmini series, UV-1800/2450/2550/2600/2700/3600 Series
  •  When used on the UVmini series, the sample compartment unit is required.

S-1700 Thermoelectric Single Cell Holder

S-1700 Thermoelectric Single Cell Holder

This cell holder enables setting of the temperature program to raise and lower the sample cell temperature. The thermoelectric system enables the sample temperature to be controlled between 0 and 110 °C. A stirrer is also provided to ensure uniform temperature distribution throughout the cell.

  • For UVmini series, UV-1800/2450/2550/2600/2700/3600 Series
  • Cooling water temperature: 20±2 °C
  • Flow rate: 4.8 L/min or more
  •  Temperature is not controlled on the reference side.
  •  Cooling water circulation is required for cooling the Peltier element. Though tap water can be used for cooling, the following conditions must be satisfied to elicit the maximum performance of the instrument.
    Use of a commercially available constant-temperature water circulator is recommended.
  •  Cells are not provided. The following 10 mm square cell with a stopper (made by Hellma) is required.
Type Optical Path Length Minimum Required Sample Volume
110-QS-10 10 mm 3.5 mL
115B-QS-10 10 mm 400 µL

NTT-2200PP Constant-Temperature Water Circulator

NTT-2200PP Constant-Temperature Water Circulator

This accessory circulates temperature-controlled water to a constant-temperature cell holder.

  • Temperature control range: 35 to 80 °C
    (ambient +5 to +80 °C at room temperature 30 °C or more)
  • Temperature control precision: ±0.05 °C or more
  • Max. pump circulation rate: 27/31 L/min
  • Max. pump head: 9.5/13 m (50/60 Hz)
  • External circulation nozzle: 10.5 mm O.D. (both outlet and return)
  • Tank capacity: Approx. 10 L (9 L during operation)
  • Standard accessories: Lid with handles, hose 4 m (I.D. 8 mm dia., O.D. 12 mm dia.), 4 hose clamps, Instruction Manual
  • Size: 270 W x 560 H x 400 D (mm)
  • Power requirements: 100 VAC, 1250 VA, with 1.7 m power cable and grounded plug

For Research Use Only. Not for use in diagnostic procedures.