LC-20AB Pump

Binary Solvent Delivery Unit

LC-20AB pump

The LC-20AB is a binary, high-pressure gradient solvent delivery unit that incorporates two LC-20AD pumps into one compact unit. The space-saving design and superior performance of the LC-20AD pump provide precise and accurate high-pressure gradients.

Features

Enhanced Solvent Delivery Specifications
Flow-rate accuracy and precision in the micro-flow rate region have been improved with the addition of a new type of check valve and modifications to the solvent delivery control method. Non-polar organic solvents such as hexane can be delivered stably.

Improved Durability
New and improved plunger seal materials ensure robust, long-lasting performance. A sapphire plunger optimized according to its crystal structure and a precise plunger-holding mechanism help increase the service life of plunger seals and enable stable solvent delivery over long periods.

Space-saving Designs
An optional low pressure gradient valve (LC-20AD/20AT), CBM-20Alite system controller, and a seal wash pump (LC-20AD) can now be installed inside the solvent delivery unit, enabling 4-solvent gradient elution in a compact space. The LC-20AB high-pressure binary gradient pump delivers the performance of two LC-20AD pumps in one compact unit.

Specifications

  LC-20AD LC-20AT LC-20AB
Solvent delivery method Parallel-type double plunger Series-type double plunger Parallel-type double plunger (2 sets)
Plunger capacity 10µL Primary side: 47µL, Secondary side: 23µL 10µL
Maximum discharge pressure 40MPa
Flow-rate setting range 0.0001 to 10.0000 mL/min. 0.001 to 10.000 mL/min. 0.0001 to 10.0000 mL/min.
Flow-rate accuracy No more than 1% or 0.5µL/min., whichever is greater (0.01 to 2 mL/min.) No more than 2% or 2µL/min., whichever is greater (0.01 to 5 mL/min.) No more than 1% or 0.5µL/min., whichever is greater (0.01 to 2 mL/min.)
Flow-rate precision No more than 0.06% RSD or 0.02 min SD, whichever is greater
Pulsation 0.1 MPa (for water at 1.0 mL/min. and 7MPa) 0.3 MPa (for water at 1.0 mL/min. and 7MPa) 0.1 MPa (for water at 1.0 mL/min. and 7MPa)
Constant-pressure solvent delivery Supported
Plunger rinsing mechanism Manual rinsing or automatic rinsing using optional product
Safety measures Liquid-leakage sensor, high-pressure/low-pressure limits
Operating temperature range 4oC to 35oC
Dimensions, weight 260 (W) x 140 (H) x 420 (D) mm, 10 kg 260 (W) x 140 (H) x 420 (D) mm, 11 kg 260 (W) x 140 (H) x 420 (D) mm, 13 kg
Power requirements 100 VAC, 150 VA, 50/60 Hz 100 VAC, 180 VA, 50/60 Hz
GRADIENT SYSTEM
  LC-20AD/20AT, low-pressure GE specifications LC-20AD/20AT, high-pressure GE specifications LC-20AB
Gradient type Low-pressure mixing High-pressure mixing
Number of mixed solvents 4 max. 2 or 3 2
Mixing concentration accuracy   0.5% max. (LC-20AD), 1.0% max. (LC-20AT) 0.5% max.

For Research Use Only. Not for use in diagnostic procedures.